Инструкция представлена в формате PDF, просмотр, скачивание на компьютер, смартфон бесплатно. Возможность печати зависит от используемой программы и настроек устройства. Содержание, структура и объём документа могут различаться в зависимости от модели, версии и редакции руководства. Перед использованием рекомендуется убедиться, что инструкция соответствует вашей модели.

Назад
/ 48
Вперёд
5.5.Inflating further
If the cuff is not inflated sufficiently, the measurement is broken off
after a few seconds and a flashing arrow pointing upwards appears
in the display. It is then necessary to inflate further to at least
20 mmHg higher than the previous value. The instruction to inflate
further can appear several times if the pressure is still not sufficient.
Attention:
If the cuff is inflated too far (over 300 mmHg), «HI» appears in the
display and a warning signal is sounded. The cuff pressure must
then be immediately released by pressing the «0/I» button.
As soon as the cuff is inflated sufficiently and the measurement is
proceeding, continue as described under points c-d.
5.6.Discontinuing a measurement
If it is necessary to interrupt a blood pressure measurement for
any reason (e.g. the patient feels unwell), the 0/I button can be
pressed at any time.
The device immediately lowers the cuff-pressure automatically.
5.7.Memory – displaying the last measurement
The measured results are stored in the instrument until
a new measurement is carried out or the batteries are
removed. You can call up the stored values at any time by
pressing the O/I button for at least 3 seconds.
6. Error messages/malfunctions
If an error occurs during a measurement, the measurement
is discontinued and a corresponding error code is displayed.
(Example Error no. 1)
* Please consult your doctor, if this or any other problem occurs repeatedly.
7
MR
Error No. Possible cause(s)
ERR 1 The systolic pressure was determined but afterwards the cuff pressure fell below 20 mmHg.
This situation can occur, for example, if the tube becomes unplugged after the systolic blood
pressure has been measured. Further possible cause: No pulse has been detected.
ERR 2 Unnatural pressure impulses influence the measurement result. Reason: The arm was
moved during the measurement (Artefact).
ERR 5 The measured results have indicated an unacceptable difference between the systolic and
diastolic pressure. Take another measurement, carefully following the instructions. Contact
your doctor, if you still obtain unusual results.
HI
The pressure in the cuff is too high (over 300 mmHg) OR the pulse is too high (over 200
beats per minute). Relax for 5 minutes and repeat the measurement.*
LO
The pulse is too low (less than 40 beats per minute). Repeat the measurement.*